[PDF] URGENT: MEDICAL DEVICE CORRECTION



Previous PDF Next PDF







AN URGENT LETTER TO ALL EMPLOYERS

AN URGENT LETTER TO ALL EMPLOYERS Dear CEO, When required, the decision to furlough normal workers on PAYE is a relatively simple one Employers have two options: 1) Keep workers on the payroll and put them on furlough leave 2) Give them notice of redundancy



INFORMATION REQUEST LETTER URGENT LEGAL MATTER-PROMPT REPLY

INFORMATION REQUEST LETTER URGENT LEGAL MATTER-PROMPT REPLY REQUESTED CERTIFIED MAIL-RETURN RECEIPT REQUESTED Katherine Lynn Director Alabama Build1ng Commission 770 Washington Avenue, Suite 444 Montgomery, Alabama 36104 Re: Request for Information Concerning Property Located at 500 & 501 Dexter Avenue,



Urgent Payment Request Letter - uploadsstrikinglycdncom

Urgent Payment Request Letter Tralatitious Radcliffe clarified her Puccini so knowledgeably that Levy redacts very impurely Salem remains contemporary after Renaud outdistanced promisingly or captivated any sing



URGENT: MEDICAL DEVICE CORRECTION

URGENT: MEDICAL DEVICE CORRECTION November 5, 2014 Dear Valued Medtronic Customer: Medtronic is committed to keeping you and your healthcare professional informed of issues and solutions concerning our products and services The purpose of this letter is to inform you of a potential safety issue



Thank you for your attention to this urgent matter, and we

May 01, 2020 · The Aluminum Association sent a letter to Commerce Secretary Wilbur Ross on April 22 nd about this issue, and we strongly urge you to address this as soon as possible Second, we believe your “phase one” deal with China was a missed opportunity



Open Letter from the UK Medical Freedom Alliance to

Re: Urgent warning re Covid-19 vaccine-related deaths in the elderly and Care Homes In our Open Letter of 23 November 2020, addressed to the MHRA, JCVI and Matt Hancock i , we outlined our concerns of potential public health risks from a mass roll-out of the Covid-19 vaccines



[DRAFT LETTER TO SCHOOL PRINCIPAL] [INSERT YOUR ADDRESS

[DRAFT LETTER TO SCHOOL PRINCIPAL] [INSERT YOUR ADDRESS] To: The Principal [INSERT address of recipient] Date: NOTICE OF LIABILITY URGENT ACTION REQUIRED RE: [INSERT NAME OF PUPIL AND THEIR YEAR/CLASS DETAILS] Dear [INSERT name/title], I am a parent of the above named pupil I writing to you to express my grave concerns regarding the



Interpretive Letter Ғ

Interpretive Letter #1170 July 22, 2020 July 2020



Welcome Letter & Guide For Our Patients

URGENT CARE WALK IN urgent care is available for all of our registered patients This service is available Monday-Friday 8:30am-10:30 am Our goal is to provide urgent medical care for acute illness within one hour of your arrival APPOINTMENTS Sheridan Medical Group is committed to providing quality care to our patients To ensure



DCS5525 URGENT - DISTRIBUTE IMMEDIATELY Date: October 1, 2020

URGENT - DISTRIBUTE IMMEDIATELY Date: October 1, 2020 Subject: N192268180 - Emission Recall Tailpipe Emissions – Oxygen Sensor Models: 2012-2013 Chevrolet Equinox 2012-2013 GMC Terrain To: All General Motors Dealers General Motors is releasing Emission Recall N192268180 today The total

[PDF] lettre victor hugo les contemplations

[PDF] Lettre: correction

[PDF] lettres ? décorer ? imprimer

[PDF] Lettrés ? un professeur d'Arthur Rimbaud

[PDF] lettres a mme de sévigné

[PDF] Lettres Anglaises de Voltaire

[PDF] lettres choisies de mme de sévigné

[PDF] Lettres choisies, Mme de Sévigné, Fleur d'Encre 4ème, hachette Epistolaire

[PDF] Lettres d'amour de 10 ? 10

[PDF] Lettres de Joseph à Louis Pasteur

[PDF] Lettres de l'intérieur de John Marsden

[PDF] lettres de madame de sévigné nuit de noces

[PDF] lettres de madame de sévigné pdf

[PDF] lettres de madame de sévigné texte intégral

[PDF] lettres de marraines de guerre 14-18

Medtronic, Inc.

18000 Devonshire St.

Northridge, CA 91325

www.medtronicdiabetes.com

Pa ge 1 of 2

*This field notification involves models MMT-368, MMT-368T, MMT-371, MMT-371T, MMT-373, MMT-373T, MMT-377, MMT-377T, MMT-378, MMT-378T, MMT-381,

MMT-381T, MMT-382, MMT-382T, MMT-383,MMT-383T, MMT-384 and MMT-384T, MMT-862, MMT-862T, MMT-863, MMT-864, MMT-864T, MMT-865, MMT-866,

MMT-866T, MMT-873, MMT-874, MMT-874T, MMT-875, MMT-876, MMT-876T, MMT-883, MMT-884, MMT-885, MMT-886, and MMT-886T infusion sets.

URGENT: MEDICAL DEVICE CORRECTION

November 5, 2014

Dear Valued Medtronic Customer: Medtronic is committed to keeping you and your healthcare professional informed of issues and solutions

concerning our products and services. The purpose of this letter is to inform you of a potential safety issue

regarding the MiniMed Silhouette and MiniMed Sure-T infusion sets*, which our records indicate you have purchased.

As part of our global quality monitoring process, we identified an increase in reports of the tubing becoming

detached at the connect/disconnect location on the MiniMed Silhouette and MiniMed Sure-T infusion sets

(please see images below). While the number of these reports remains low, you are being notified of this

potential issue because your safety is our top priority.

If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify you. The

interruption of insulin delivery can cause hyperglycemia, which if left untreated, can result in diabetic

ketoacidosis (DKA). DKA is a serious condition that can cause a severe impact to health, including death. Symptoms of DKA may include nausea, vomiting, shortness of breath and excess thirst/urination. Seek

medical attention immediately if you are experiencing any of these symptoms.

What action do I need to take?

You can continue to use your infusion sets by following the advice below:

1.When chan ging your infusion set, closely follow the instructions for use included in the product box.

Check the tubing at the connect/disconnect location identified in the drawings above to make sure it is

not loose.

2.As always, it is essential to monitor your blood sugar levels frequently using your blood glucose

meter. Proactively check your tubing connections occasionally throughout the day to ensure tubing is secure. It is especially important to check your blood sugar and tubing connections at bedtime to confirm insulin delivery is occurring.

3.If you experience a hi gh blood sugar, check your tubing connections and infusion site closely to

ensure your tubing is secure.

Example of Tubing Detachment

930M12843-011 20140905

© 2014 Medtronic MiniMed, Inc. All rights reserved. MiniMed, Silhouette, Sure-T, Quick-set and Mio are trademarks of Medtronic MiniMed, Inc.

Page 2 of 2

If you discover the tubing is detached:

1.Do not att empt to reattach the tubing. Replace the infusion set immediately.

2.Treat any high blood suga r based on guidelines provided by your healthcare professional.

3.Call th e Medtronic 24-Hour HelpLine at 866.222.7304 to report the issue. We will give you instructions

on how to return the affected infusion set to us.

We have identified a solution and are working to incorporate improvements within the coming weeks. In the

meantime, you will still be able to order these infusion sets. If you would like to try another type of infusion set

instead, we can exchange your infusion sets with the MiniMed Quick-set or MiniMed Mio . To learn more

about these infusion sets or to find additional information about this notification, please call us or visit

You can always report a concern to th

e FDAÕs MedWatch Adverse Event Reporting program: Online at: http//www.fda.gov/safety/medwatch/howtoreport/default.htm

Report by telephone: 1.800.FDA.1088

Fax report: 1.800.FDA.0178

We appreciate your time and attention to this important notification.

Sincerely,

Jim Dabbs

Vice President, Quality Assurance

Medtronic Diabetes