echelle spectrometer that uses the latest developments in camera technology and instrument manufacturing The result is unmatched performance and robustness at an entirely new price level Meet IRIS, the next generation echelle spectrometer Introduction A typical definition of an echelle spectrometer is a



Echelle Optics Explained Simply - Horiba

As mentioned earlier, the echelle uses multiple orders and wavelength-efficiency curves can be described as “order hopping”, as shown in Figure 4 These echelle characteristics give signal “holes”, just before and just after the order hop, where the signal intensity is low In the UV region, this order hopping may occur every 4 nm



Fundamentals of Echelle Spectroscopy - Vilniaus universitetas

Fundamentals of echelle spectroscopy Moletai Summer School ­ August 2012 Example ­ possible resolution for FIES FIES has an echelle with 31 6 grooves/mm and a blaze angle of 63 5 degrees The FIES beam has a diameter of 15 cm The NOT mirror has a diameter of 250 cm



Avio 550/560 Max ICP-OES Optical System and SCD Detector

array charge-coupled device (SCD) detector An echelle-based polychromator was designed to fully utilize the capabilities of the SCD Using a PerkinElmer echelle grating optimized for UV performance and a free-form optic, the Schmidt cross-disperser, the Avio 550/560 Max ICP-OES systems have exceptional optical throughput and excellent



Echelle d’évaluation des Répercussions Fonctionnelles (ERF)

Echelle ERF P Vianin, 2014 COTATION La cotation se fait en fonction de deux éléments (cf éléments de cotation, p 8 - 12) : - le contenu de la pensée exprimée par le patient au cours de l’entretien, c’est-à-dire les



Etapes de la classification Echelle de déficience ASIA

Echelle de déficience ASIA pas la motricité est préservée au dessous Etapes de la classification 1 Déterminer les niveaux sensitifs pour les côtés droit et gauche Le niveau sensitif est le dernier dermatome sain pour la piqûre et le toucher 2 Déterminer les niveaux moteurs pour les côtés droit et gauche



Echelle de Clance du Phénomène de l’Imposteur

Echelle de Clance du Phénomène de l’Imposteur Pour chaque question, entourer le nombre qui indique le mieux à quel point la phrase est vraie pour vous Il vaut mieux donner la première réponse qui vous passe par la tête et ne pas trop réfléchir à chaque phrase 1



Les accessibilités Voie engin Voie échelle

Version du 14 octobre 2013 14/16 Les contrôles principaux Accès voie engin conforme en emprise (stationnement) et en giration (carrefours )

[PDF] echelle 1/250 sur un plan

[PDF] echelle 1/250 conversion

[PDF] echelle 1/25000

[PDF] formule effet mix controle de gestion

[PDF] analyse des écarts controle de gestion

[PDF] effet taux effet volume

[PDF] calcul cout standard

[PDF] convertir en base 5

[PDF] passer d'une base ? une autre

[PDF] conversion binaire decimal hexadecimal

[PDF] algorithme de conversion de base

[PDF] conversion base 16 en base 2

[PDF] convertir en base 8

[PDF] calculer avec des lettres

[PDF] lecon calcul litteral 4ème

Echelle Optics Explained SimplyGeoff Tyler

Jobin Yvon S.A.S., Horiba Group

Longjumeau, France

Keywords: optics, echelle, orders

A question often asked is "What is an echelle

spectrometer? Why are they used?" Originally invented for astronomy, echelle optics were first used in atomic spectroscopy on the DCP (direct current plasma) and more recently in ICP (induc- tively coupled plasma). It is a very compact opti- cal design of less than 0.4 meters long, incorpo- rating usually a course grating e.g. 120 g/mm and a prism as a cross disperser. Instead of hav- ing a linear spectrum, with intensity of line against wavelength, the echelle, using a cross disperser, creates a two dimensional spectrum with wavelengths across and in a series of lines going down, like the text in a page of a book.

Figure 1 shows a typical echelle spectrum.

Figure 1: Echelle two dimensional spectrum

There are many designs of echelle optics, most

use prisms, (some designs use a grating as the cross disperser), the basic format however, is similar. The course grating creates low resolu- tion, then the cross disperser creates enhanced resolution two dimensionally. The coarse grating and the prism utilize multiple orders, typically between 50-120, to cover the normal spectral range required for ICP.This optical design is ideal for simultaneous solid state detector (SSD) ICP spectrometers due to the shape and size of the chip. The shape of the spectrum is an off-axis bell shape (Figure 2), however, about 30% of the chip area is not used.

There are also limitations on the wavelengths

available, due to the focusing optics, the chip size, the number of pixels per chip, and the desire to use as much of the chip area as possible, with minimum loss of wavelengths falling off the edge.

Figure 2: Echellogram: Off-axis echelle spectrum

With any optical system the use of higher orders

will provide better resolution, smaller wavelength range (for that order) and an increased back- ground due to stray light. This phenomenon is especially found with multiple order systems.

Increased background causes more noise and

poorer precision, especially with low signals as observed with low concentrations. The stray light of echelle optics is especially high due to the very high orders used. This in turn affects the detec- tion limit.ICP OPTICAL EMISSION SPECTROSCOPY

TECHNICAL NOTE 04

2

ICP OPTICAL EMISSION SPECTROSCOPY

TECHNICAL NOTE 04

The two dimensional echelle spectrum has a vari-

able resolution that can be fairly good in the UV region. However, one characteristic of echelle optics is that the resolution gradually changes with wavelength. The prism usually used produces the highest resolution in the UV region but for the lower orders (higher wavelengths) they have much poor- er resolution, sometimes, almost a factor of 10 times worse. Resolution is accepted by all to be very important in

ICP-OES. However, terms like "pixel" resolution,

when used with SSD spectrometers can often raise confusion compared to the classical definition of full width at half height (FWHH) or peak valley methods. In SSD spectrometers using echelle optics, a peak will typically be characterized by 3-4 pixels, so for example, a 7 pm "pixel" resolution spectrometer might yield twice that figure when using the FWHH definition of optical resolution.

This can be seen in classical resolution tests of

As/Cd at 228.802 nm (Figure 3). Mathematical cor-

rections can improve detection limits in some cases, but usually it corrects for problems created by opti- cal deficiencies.

Figure 3: As/Cd with 5pm optical resolution as

acquired on a JY ULTIMA 2 ICP-OES. Another characteristic of echelle optics is that an aberration occurs causing a blurred image on the edge of the bell shaped spectrum. This leads to a light loss (less light per pixel) and poor resolution compared with wavelengths in the center. Therefore, with these echelle spectrometers a reso- lution of 10 pm can degrade to 25 pm at the edge, even though it may be less than 2 nm away. This phenomenon will show as poor detection limits and resolution for that particular elemental wavelength. Even on classical optics the resolution can change as the wavelength range limit is reached. For high resolution, either a very high-density grating is used e.g. 4320 g/mm, or several grating orders are used with order sorting by means of a solar blind PMT or filters. Either option gives a restricted wavelength range for a given resolution. For example, a 4320 g/mm grating will have a range of 120-450 nm on

1 meter optics. A 2400 g/mm grating on the same

optical system using 2ndorder will cover 120-320 nm, while 1storder will cover the high wavelengths of 120-800 nm.

As mentioned earlier, the echelle uses multiple

orders and wavelength-efficiency curves can be described as "order hopping", as shown in Figure 4. These echelle characteristics give signal "holes", just before and just after the order hop, where the signal intensity is low. In the UV region, this order hopping may occur every 4 nm. 3

ICP OPTICAL EMISSION SPECTROSCOPY

TECHNICAL NOTE 04

Figure 4: Echelle optics "order hopping" spectrum

from a constant light source. Note the degraded resolution with wavelength, and the amplitude of the intensity peak to valley. Due to order hopping, the slit height of the echelle spectrometer must be small in comparison with classical optics. If a normal slit were used e.g. 6-8 mm, the spectrum from the various orders would merge together giving spectral overlap and gross interferences. For this reason an echelle SSD spec- trometer will typically use 0.2 mm slit height. Even though an echelle spectrometer can use wider slits than classical optics, the difference in slit height leads to a large difference in total light throughput to the detector between the two optical systems. Figure 5 shows a comparison of energy at different slit heights.Figure 5: Slit height and energy comparison

In the USA:

Jobin Yvon Inc.

3880 Park Avenue

Edison, NJ 08820

Tel: 1-732-494-8660

Fax: 1-732-494-8796

1-877-JYHORIBA

In France:

Jobin Yvon S.A.S.

16-18, rue du Canal

91165 Longjumeau Cedex

Tel: (33) 1/64 54 13 00

Fax: (33) 1/69 09 90 88

Germany: (49) 89/46 23 17-0

Italy: (39) 2/57 60 56 90

U.K.: (44) 20/82 04 81 42In Japan:

Horiba Ltd.

2 Miyanohigashi, Kisshoin

Minami-ku, Kyoto 601-8510

TEL: (81) 75 313 8121

FAX: (81) 75 321 5725

www.horiba.com

China: (86) 10/6836 6542

Spain: (34) 91/724 16 57

Other Countries: Contact JY

S.A.S.

quotesdbs_dbs5.pdfusesText_10